HP-480 là thiế bị đo pH loại gắn bảng do HORIBA sản xuất.
Mô tả: HP-480 là máy đo chất lượng nước nhỏ gọn cung cấp nhiều tính năng và chức năng. Nó phù hợp để theo dõi và kiểm soát độ pH trong các quy trình sản xuất khác nhau, cũng như xử l...
| Product name |
Industrial pH meter |
| Model |
HP-480 |
| Measuring range |
0 pH to 14 pH: 0.01 pH resolution.
Temperature 0℃ to 100℃: 1℃ resolution (selectable display). |
| Transmission
output |
4 mA to 20 mA DC: input/output isolated type
Maximum load resistance: 900 Ω |
| Transmission
output range |
Free |
| Repeatability and
linearity |
±0.05 pH, ±0.08 mA (for transmission output range of 0 pH to 14 pH) |
| Contact output |
Outputs: 2 points
Alarm contact output (R1 and R2)
Contact type: relay contact, SPDT (1c)
Contact rating: 240 V AC, 3 A and 30 V DC, 3 A (resistance load)
Contact function: selectable from upper/lower limit operation (ON/OFF control), alarm, and maintenance. |
| Calibration function |
Two-point automatic calibration or manual calibration.
Two-point automatic calibration: with automatic validation of the electrode characteristics.
Types of standard solution: pH 2, pH 4, pH 7, pH 9 and pH 10 (JIS)
Combination of standard solutions: pH 7 and one of the others
Manual calibration: Freely selectable, but the difference should be over 2 pH.
Temperature calibration (one point). |
| Transmission
output hold feature |
- Previous value hold
- Optional value hold
- Continuous
Selectable from the above (However, only the previous value hold is available in the maintenance mode).
|
| Self-diagnosis
function |
- Calibration function
Asymmetry potential error, sensitivity error, response speed error and standard solution error.
- Electrode self-diagnosis
Temperature sensor short-circuit and temperature sensor disconnection.
- Outside of the measuring range.
- Converter error.
|
| Temperature
compensation
element |
Selectable from compensation ON (350Ω (25℃), 500Ω (25℃),1 kΩ (0℃)6.8 kΩ (25℃)
and 10 kΩ (25℃)) and compensation OFF. |
| Temperature-
compensated
range |
0℃ to 100℃ |
| Ambient
temperature |
−5℃ to 45℃ |
| Relative humidity |
20% to 85% (without dew condensation) |
| Storage
temperature |
−25℃ to 65℃ |
| Power supply |
Rated voltage 100 V AC to 240 V AC, 50 Hz/60 Hz, 10 VA (max.) |
| Structure |
Indoor-use panel installation type
Panel case: ABS, Terminal: PBT
Panel: water proof structure |
| Protective structure |
Panel: IP65 (IEC60529, JIS C0920)
Rear case: IP20, Terminal: IP00
Class II device (IEC61010-1)
Pollution level 2 (IEC61010-1) |
| Conforming
standards |
CE Marking |
EMC Directive: EN61326-1
Low Voltage Directive: EN61010-1
RoHS Directive: EN50581 |
Immunity (Industrial electromagnetic environment)
Noise increase ±0.2 pH or less |
| FCC Rule |
FCC Part15 |
| External
dimensions |
48 (W) × 96 (H) × 115 (D) mm
Case depth: approx. 105 mm |
| Mass |
Approx. 400 g |